HOME Products Back

RBRquadrante four-channel radiometer

Product Number:RBRquadrante four-channel radiometer
The RBRquadrante is a multi-spectral radiometer with four channels, capable of measuring multiple wavebands simultaneously, including PAR. It features a high dynamic range, optimized cosine response, and excellent low-light detection, while power consumption and depth rating have been tailored for use in a wide variety of applications.

The RBRquadrante supports measurement of four wavebands within the same sensor package. Tolerant of a wide ranging power supply, data are streamed via RS-232 on the MCBH-6-MP connector. A variety of adaptors make it easy to integrate into existing vehicle payload bays.
  • Low power consumption
  • High accuracy
  • High dynamic range
  • Depths up to 2000m
  • RS-232 output
  • Compact and lightweight
Physical
Connector MCBH-6-MP
Diffuser Acrylic
Housing Titanium
Diameter 63mm
Length 57mm, 93mm (with connector)
Depth rating 2000m
Weight (air) 400g (in air), 210g (in water)
Sampling rate Up to 32 Hz
Power consumption
Supply voltage 4.5 to 30V (12V nominal)
Sampling 4mJ/sample (4Hz or slower)
3mA/36mW (8Hz or faster)
Sleep current 10μA
Optical radiometry
Dynamic range >5.5 decades
Absolute calibration±5%
Linearity ±1%
Operating temperature range -5°C to 35°C
Cosine response error (water) ±5% at 0-60°
±10% at 61-82°
Azimuth error (water) ±1.5% at 45°
Out-of-band rejection>25dB (typical), OD 2.5
Photosynthetically active radiation
Wavelength range 400 to 700nm
Full scale range 0-5000µmol/m²/s (minimum)
Resolution ±0.010μmol/m²/s
Narrow-band wavelength channels
Centre wavelengths (CWL) 413 / 445 / 475 / 488 / 508 / 532 / 560nm
Full width at half-maximum (FWHM) 10nm (for all CWLs except 475nm)
25nm (for CWL 475nm only)
Full scale range 0-400μW/cm2/nm (minimum)
Resolution3 ±0.001μW/cm2/nm
1RBR calibrates radiometers with NIST traceable references.
2Out-of-band rejection and resolution are wavelength dependent for narrow-band radiometers.
3Resolution is wavelength dependent for narrow-band radiometers.
Note: Dark offset is internally temperature-compensated.