HOME Products Back

RBRduo³ and RBRconcerto³: C.T, C.T.D, C.T.D++

Product Number:RBRduo³ C.T and RBRconcerto³ C.T.D
Our flexible, option-packed loggers dedicated to salinity and more.

240M readings
Standard sampling of 2Hz
Up to 32Hz sampling
USB-C download
Realtime communications
Twist activation
Wi-Fi communication option

C.T conductivity, temperature sensors
C.T.D conductivity, temperature, pressure (depth) sensors
C.T.D + C.T.D plus one additional sensor
C.T.D ++ C.T.D plus two additional sensors

Available sensors:

Conductivity (C), Temperature (T), Pressure (D), Dissolved Oxygen (DO), Optical Dissolved Oxygen (ODO), Photosynthetically Active Radiation (PAR), Radiometer (rad), Turbidity (Tu), Thermistor string (Tx), Fluorescence (Fl), pH, Oxidation-Reduction Potential (ORP), Methane, Carbon Dioxide, Transmittance, Voltage
RBRduo³ C.T and RBRconcerto³ C.T.D
The RBRduo³ C.T and RBRconcerto³ C.T.D are built on the years of experience RBR has creating reliable and easy-to-use instruments. They are uniquely designed to determine salinity by measuring the conductivity and temperature of water. Equipped with a pressure channel, the RBRconcerto³ C.T.D can also derive depth, density anomaly, and speed of sound. These self-contained, accurate, and autonomous submersible instruments can be deployed for a period of up to five years at 1min sampling intervals, and the data is retained in memory even after the batteries are removed.

Salinity
Conductivity measurements are performed using a rugged inductive cell suitable for deployment in marine, estuarine, or fresh water. There are no exposed contacts, which avoids susceptibility to corrosion, and the housing may be frozen into ice without damage. It is characterized for the mechanical effects of both temperature and pressure to allow for accurate correction in salinity derivation. These coefficients are provided for customers who prefer to calculate their own corrections. In addition, it is suitable for use outside the usual realms of PSS-78 practical salinity scale, in applications such as desalination brine monitoring. The conductivity readings are temperature-corrected.

For profiling applications, the co-located thermistor reduces salinity spiking, thus increasing data accuracy. 

Equipped with a piezoresistive pressure channel, the CTD instruments provide more accurate salinity data when the instrument is sampling at varying depths.  

Additional sensors
A diversity of sensor configurations allows the instrument to be fine-tuned for a wide variety of applications. Variants with pressure, temperature, conductivity, radiometer, PAR, and turbidity sensors are also available in titanium housing, designed to endure harsh conditions.
Physical
Storage 240M readings
Power1 8 AA cells
External power 4.5 to 30V
Communication USB-C or RS-232/485
Clock drift ±60 seconds/year
Housing Plastic or titanium
Diameter 63.3mm (plastic), 60.3mm (Ti)
Length Configuration dependent
Weight Configuration dependent
Depth ratingUp to 6000m (configuration dependent)
Sampling rate 2Hz; options up to 32Hz

1 Lithium thionyl chloride batteries are only recommended for the RBRduo³ C.T and RBRconcerto³ C.T.D. Use alkaline or lithium iron batteries for all other configurations.
2 The depth rating for RBRconcerto³ Tx may be up to 8000m.
Conductivity
Range 0-85mS/cm
Initial accuracy ±0.003mS/cm
Typical stability 0.010mS/cm/year
Resolution 0.001mS/cm
Temperature
Range-5°C to 35°C
Initial accuracy ±0.002°C
Resolution <0.00005°C
Time constant <0.1s | fast, <1s standard
Typical stability 0.002°C/year

3A wider temperature range is available upon request. Contact RBR for more information.
Pressure (Depth)
Range20 / 50 / 100 / 200 / 500 / 750dbar (plastic)
/1000 / 2000 / 4000 / 6000dbar (Ti)
Initial accuracy ±0.05% full scale
Resolution <0.001% full scale
Time constant <0.01s
Typical stability ±0.05% full scale per year

4Recommended depth for wave measurements is less than 50m.
Turbidity (Seapoint®)
Light source 880nm
Time Constant 0.1s
Linearity <2% deviation for 0-1250 FTU
Fluorescence (Seapoint®)
Excitation 470nm / 685nm (chlorophyll a)
370nm / 440nm (cDOM)
Time Constant 0.1s
Dissolved Oxygen Sensor (Oxyguard®)
Range 0 to 600%
Initial accuracy ±2% O2 saturation
Resolution 1% of saturation
Response Time ~10s, 90% step change at 20°C
Contact RBR for other sensor options.

Inquriy Cart

Total 0 items

    Search..

    Please Enable cookies to improve your user experience